Biofluorescent particle counters: Detecting cleanroom contamination

BY DR TIM SANDLE | PHARMACEUTICAL MICROBIOLOGY AND CONTAMINATION CONTROL EXPERT

19th December

 

In a cleanroom environment, most microorganisms are not present as free-floating cells but are instead attached to larger particles such as dust, skin, or fibres, typically ranging from 1 to 40 µm in diameter1. As a result, particles measuring ≥5.0 µm are of particular interest when assessing microbial contamination, as they are more likely to act as carriers of microorganisms. Additionally, larger particles have a greater propensity to settle onto horizontal surfaces, increasing the potential risk of surface contamination.

 

Determining the extent to which these larger particle rafts carry microorganisms is challenging. The limitations of culture-based monitoring methods are well recognised, with a conventional light-scattering airborne particle counter detecting particles only, without providing any direct indication of the presence of viable organisms 2.


Biofluorescent particle counters

 

Alternative technologies for microbial detection are available. First marketed around 2010, these systems can differentiate between inert and biologic particles. Known as biofluorescent (or spectrophotometric) particle counters, these devices provide real-time indications of probable airborne microbial contamination. The term ‘probable’ is used advisedly, as these instruments can be susceptible to false positives. 


In addition, there is no direct quantification from a detected event with actual microbial cells.
Nevertheless, biofluorescent particle counters can serve as effective early warning tools, signalling potential out-of-control environmental situations and enabling timely corrective action.

 

 

What is Biofluorescence?

Biofluorescence is a type of fluorescence exhibited by a living organism. Here, part of the organism absorbs light (such as blue light or

ultraviolet light) at one wavelength and emits visible light at a longer wavelength (such as green or red light).

 

The term should not be confused with bioluminescence.

 

 

In terms of their operation, these devices analyse laser-induced fluorescence pulses emitted by airborne microbiological particles as they are drawn into the counter. Internal detectors are tuned to identify biologic entities, specifically the fluorophores tryptophan, reduced Nicotinamide Adenine Dinucleotide (NADH) and riboflavin (detected using a blue laser with 405-nm wavelength as its excitation source). These metabolites are commonly found fluorophores in bacterial cells 3

 

The instrument stimulates and then collects the total fluorescence signal from each probed aerosol particle captured, allowing the determination of the ratio of elastic scattered light at the laser wavelength to fluoresced light. Simultaneously, the device will allow particle sizing - much like a conventional particle counter - through the collection of the elastically scattered laser light and a flux measurement, by counting the particles that pass 4.

 

 

Advantages

 

These counters have several uses and connect with the Process Analytical Technology (PAT) paradigm. One use is the monitoring of aseptic processing environments, enabling filling runs to be stopped if biologic events are detected. Having a ‘real time’ capability eliminates the inherent delays of culture-based methods 5. The counters can provide improved risk-reduction as well as providing a rapid indication of the success (or otherwise) from root cause analysis and in putting in place remedial actions 6.

 

Moreover, the counters provide a higher estimation of contamination, helping to overcome the limitations of culture-practices. Since culturing the organisms is not necessary for detection, biofluorescent counters can detect Viable but Non-Culturable (VBNC) microorganisms.

 

 

Limitations

 

Biofluorescent particle counters do have limitations. One limitation is false positives, where the fluorophores present in artificial products can interfere with airborne assessments, leading to the incorrect detection of microorganisms when these materials are present as airborne particles in a cleanroom. Fluorophores can arise from clothing fibres, spray-droplets of disinfectants or as aerosolised debris made of fluorinated rubber product 7, 8.

 

Manufacturers have sought to limit these false positives, with technologies advancing since their inception. The verification of biofluorescent counters can be improved using inkjet aerosol generators, where the use of two types of monodisperse test particles - inert particles and fluorescent particles - can be used to assess the potential for false positives and enable better tuning to eliminate the rate of false occurrences 9

 

 

Challenges: Colony Forming Units (CFUs)

 

Another issue with biofluorescent particle counters is how the data recorded for a suspected biologic event translates to actual microbial cells. The reported unit is commonly referred to as an Autofluorescence Unit (AFU), which is not dependent upon growth, as is the traditional method 10. Traditional methods are reliant on counting visible (by humans) or subvisible (through blue light laser excitation) colonies.

 

It is not an easy task to make a comparison and hence validating non-growth based alternative methods, compared to the growth-based compendial method, can be fraught with difficulty (some microbiologists advise against even attempting to make a comparison) 11.

 

The reason for the discrepancy relates to the dependency on growth on agar plates so that CFUs can be counted. Since not all microbes will grow on agar and considering the VNBC issue above, standard methods for incubation do not support the growth for all microbes. For this reason, the measured values of biofluorescent systems are typically much higher than the values measured by conventional microbial tests 12.

 

Experimental attempts to correlate AFUs to the traditional CFUs are not necessarily appropriate and there is no consistent correlation factor between the two units of measure 13. A negligible to low correlation between biofluorescent particles and CFU has been reported. In contrast, a strong correlation exists with conventional particle counter measurements for the total numbers of particles 14

 

 

Applications

 

With such limitations accepted, biofluorescent particle counting can be used in other applications beyond monitoring aseptic processing environments. Examples include assessing any transfer of contamination from airlocks into higher graded cleanrooms and with setting the maximum numbers of permitted personnel in changing rooms 15. The technology can also be used to assess how long cleanrooms take to recover after facility shutdowns or unplanned events like power cuts.

 

 

Conclusion

 

Biofluorescent particle counters have been commercially available to cleanroom users for over 15 years. Their cost remains high and this is undoubtedly a reason why small to medium pharmaceutical companies cannot afford them. Those companies that have selected them tend to use them for investigational purposes and for determining facility recovery times. On the issue of false positives, as with any environmental monitoring system, the strength is with gathering as much data as possible and trending it.

 

Need expert support for your Biopharmaceutical products?

 

We have analytical experience across the entire life cycle, from early discovery to regulatory submission support. Find out more about our services and get in touch

 

References

 

1.    Malli Mohan, G., Stricker, M.  and Venkateswaran, K. 2019. Microscopic characterization of biological and inert particles associated with spacecraft assembly cleanroom. Sci. Rep. 9 (1):14251

 
2.    Sandle, T, Leavy, C, Jindal, H, Rhodes, R. Application of rapid microbiological methods for the risk assessment of controlled biopharmaceutical environments. J Appl Microbiol 2014;116:1495–1505


3.    Hill, S. Mayo, M., and Chang, R, 2009. Fluorescence of bacteria, pollens, and naturally occurring airborne particles: Excitation/emission spectra. Adelphi, MD: Army Research Laboratory


4.    Kaye, P., Barton, J., Hirst, E. and Clark, J. Simultaneous light scattering and intrinsic fluorescence measurement for the classification of airborne particles, Applied Optics, 39, 3738-3745, 2000


5.    Sandle, T. Real-time counting of airborne particles and microorganisms: a new technological wave? Clean Air Containment Rev 2012;9:4–6


6.    Prasad, A., Villari, P., Henry, R. et al. Practical Applications of Biofluorescent Particle Counting in Environmental Monitoring Investigations, PDA Journal of Pharmaceutical Science and Technology May 2020, 74 (3) 318-323


7.    Eaton, T.; Davenport, C.; Whyte, W. Airborne microbial monitoring in an operational cleanroom using an instantaneous detection system and high efficiency microbiological samplers. Eur. J. Parenter. Pharm. Sci. 2012, 17, 61–69


8.    Crawford, I., Topping, D., Gallagher, M. et al. 2020. Detection of airborne biological particles in indoor air using a real-time advanced morphological parameter UV-LIF spectrometer and gradient boosting ensemble decision tree classifiers. Atmosphere 11 (10):1039


9.    Iida, K., Ikeda, T., Minakami, T., and Sakurai, H. (2024) Verifying the viable particle counts of biofluorescent particle counters by using inkjet aerosol generators. Aerosol Science and Technology, 58(5), 554–568

 

10.    Scott, A., Vanbroekhoven, A., Joossen, C. et al. Challenges Encountered in the Implementation of Bio-Fluorescent Particle Counting Systems as a Routine Microbial Monitoring Tool, PDA Journal of Pharmaceutical Science and Technology January 2023, 77 (1) 2-9

 

11.    Martindale, C., Dreyer, C., Joossen, C. et al. Considerations for the Validation of Non-CFU Based Bio-Fluorescent Particle Counting Technologies, PDA Journal of Pharmaceutical Science and Technology August 2025, pdajpst.2024-003036.1; DOI: https://doi.org/10.5731/pdajpst.2024-003036.1

 

12.    Ayers, F., Chen, J.-P., Dingle, M. et al. Biofluorescent particle counter-based real-time feedback and control of processing conditions. Eur. Pharm. Rev. 2019, 24, 54–57

 

13.    Salvas, J., Merker, P. Dingle, M. et al. Understanding the Non-Equivalency of Bio-Fluorescent Particle Counts versus the Colony-Forming Unit, PDA Journal of Pharmaceutical Science and Technology November 2023, 77 (6) 514-518

 

14.    Stålfelt F, Caous S., J, Malchau S. et al. Real-time biofluorescent particle counting compared to conventional air sampling for monitoring airborne contamination in orthopedic implant surgery. Antimicrobial Stewardship & Healthcare Epidemiology. 2025;5(1):e93

 

15.    Sandle, T.; Leavy, C.; Rhodes, R. Assessing airborne contamination using a novel rapid microbiological method. Eur. J. Parenter. Pharm. Sci. 2014, 19, 131–141

Need to contact us? Get help or ask us a question