What is an autoclave and how does it work?

BY DR TIM SANDLE | PHARMACEUTICAL MICROBIOLOGY AND CONTAMINATION CONTROL EXPERT
31st March

 

Sterilisation by moist heat has a wide application in pharmaceuticals and healthcare. Examples of application include sterilising glassware, tools, stoppers, filters, gloves, vials, medical, laboratory and pharmaceutical items.

 

Autoclaves are steam based sterilisation devices that use moist heat under pressure to kill microorganisms, including highly resistant bacterial spores. Their operation is rooted in physics - raising pressure increases the boiling point of water, allowing steam to reach temperatures far above 100 °C (typically 121–134 °C), which is required for enhanced microbial kill and therefore for sterilisation.

 

This article examines the basic operational principles and practicalities, as required for a pharmaceutical, healthcare, medical device or food setting.

Core principle: Steam + pressure = high temperature sterilisation

 

Water normally boils at 100°C at normal atmospheric pressure. However, if pressure is increased sufficiently (e.g., +1 bar / 15 psi /100 kPa), water boils at ~12 °C. This elevated temperature creates pressurised saturated steam, which is lethal to most microorganisms, including bacterial spores likely to be found in pharmaceutical facilities.

 

In terms of the application, steam is far more efficient at transferring heat than dry air because condensation releases large amounts of latent heat directly onto surfaces, as set out in ISO 17665:2024 1

Scientist using an autoclave

There are several factors that make an autoclave effective in terms of achieving microbial kill robustly and within a sufficient period of time, whilst avoiding damaging the items to be sterilised. These factors are:

 

  • Steam penetration: Steam must contact every surface. Air pockets prevent sterilisation. This can arise from poorly defined loads. Porous items (such as garments, filters, stoppers, tubing, buckets, lumened objects) can trap air, preventing saturated steam penetration.

 

  • Heat transfer: Condensing steam transfers seven times more energy compared with dry air at similar temperatures.

 

  • Time and temperature control: Computerised systems precisely control and record cycle parameters.

Types of autoclaves

 

There are different types of autoclaves, based on their operating principles. The most common to pharmaceuticals and healthcare facilities are:

 

  • Gravity displacement autoclaves: Steam enters from top/sides and pushes air downward out of the chamber through a drain. Suitable for liquids, simple instruments, non porous loads. These are sometimes referred to as Class N autoclaves. The ‘N’ stands for naked solid products.

 

  • Pre vacuum (high speed) autoclaves: A vacuum pump is used to remove air before steam entry, improving steam penetration into porous loads, wrapped items or hollow instruments. These are sometimes referred to as Class B autoclaves - these are the highest standard. The ‘B’ represents ‘big’.

 

  • Autoclaves that combine gravity air displacement methods with limited vacuum systems. These are mid-way between N and B class and are sometimes referred to as Class S (with the S representing ‘special’). While Class S autoclaves offer greater versatility than Class N models, they do not provide the same level of effectiveness for complex materials as Class B autoclaves.

 

These autoclaves differ primarily in their sterilisation capabilities and air removal methods. The classes are defined by the European Standard EN 13060: 2025 2 (‘Small steam sterilizers’). In summary:

 

Class  Method Load Type Best Use Case
B Fractionated vacuum All (wrapped, porous, hollow) Hospitals, complex dental tools
S Partial/no vacuum Specific (manufacturer defined) Basic wrapped, solid tools
N Gravity displacement Simple (unwrapped, solid) Basic, non-hollow tools

 

The standard EN 285: 2014 3 applies to large steam sterilisers (≥ 54 litres or a full-sized sterilisation unit). The standard itself does not classify machines into types (like ‘Class B’), whereas it defines the performance, design and testing requirements for these large, usually pre-vacuum, autoclaves (in essence Class B).

 

 

How an autoclave cycle works: Step by step

 

While there are differences with autoclave design, there are some common factors which apply. These are set out below:

 

Typical example of an autoclave chart

Figure 1: A typical autoclave chart showing temperature and pressure rising through conditioning, plateauing during sterilisation and dropping during exhaust (image designed by Tim Sandle)

 

(1) Loading

  • Items must be cleaned and arranged to allow free steam circulation
  • Wrapped or hollow instruments require careful orientation
  • Use chemical and biological indicators where required

 

(2) Chamber sealing

  • The door locks with pressure safety interlocks. These seal the autoclave chamber, meaning the door cannot open when the chamber is pressurised. A heat resistant gasket ensures a steam tight seal is in place

 

(3) Air removal (critical step)

  • Air acts as an insulator, preventing steam from reaching sterilisation temperature - therefore, air must be removed
  • The mode of air removal depends on the autoclave type. Gravity units push air out the drain, while pre vacuum units extract air with a vacuum pump

 

(4) Steam admission and heating

  • Steam is introduced until the chamber reaches the programmed temperature and pressure (e.g., 121 °C or 134 °C)
  • Steam condenses on cooler surfaces, releasing high energy directly onto microbial cells, causing protein denaturation and cell death

 

(5) Sterilisation hold (exposure phase)

Depending on load type and mass, common validated conditions for the hold stage are:

  • 121 °C for 15-20 min
  • 134 °C for 3-5 min 

 

(6) Exhaust / depressurisation

  • Once the hold time has elapsed, the autoclave needs to be rendered safe for users. This requires steam to be vented safely as the pressure drops to atmospheric level

 

(7) Drying

  • The drying stage is especially important for wrapped sets to avoid ‘wet loads’ (which are a sign of non-sterility). This arises because moisture insulates the load, prevents achieving temperature and creates an environment for microbial growth. Wet load issues are also caused by poor loading configurations

 

 

 

Key performance and technical requirements

 

Under EN 285, the important performance and technical requirements are:

 

  • Steriliser applicability: This defines what the autoclave is to be used for and what is the appropriate type

 

  • Scope: This covers mechanical construction, piping, control systems, safety and performance, but generally excludes sterilisers for flammable substances or bio-waste

 

  • Steam quality tests (see Figure 2): This sets out the requirements for steam quality, including:
    • Non-condensable gases: Limit ≤ 3.5% v/v. Non‑condensable gases are gases present within steam that do not liquefy (condense) under the temperature and pressure conditions used in steam sterilisation. Instead of turning into water when steam condenses, they remain as gas pockets such as air, carbon dioxide, nitrogen, hydrogen or other trace gases. Excessive levels can lead to non-sterility (through block steam penetration, reduced heat transfer, the creation of cold spots), even where biological indicators pass
    • Dryness value: Limit ≥ 0.95. With steam dryness, steam must be saturated and collapse on contact to transfer heat. Excess wetness causes wet loads, poor dryness fraction (<97%) prevents lethality
    • Superheat: Limit  ≤25.0°C. Superheat is the temperature of steam above its saturation point. Once steam has reached the temperature at which it should condense at a given pressure (its saturation temperature), any additional temperature above is called ‘superheat’. Too much superheat means the steam is too dry, which reduces heat transfer efficiency. Steam must condense on the load to release latent heat (the lethal step)

 

Three key factors of sterilisation steam

Figure 2: Steam quality (designed by Tim Sandle)

 

 

 

Validation and testing

 

The validation and testing of autoclaves require specific, rigorous tests to be conducted to ensure sterilisation efficacy, including:

 

  • Thermometric tests: These range from small and full load tests on standard towel packs
  • Air detector tests: This ensures proper air removal
  • Air leakage tests: Limit ≤ 10-11 mbar/min
  • Equilibration time: For chambers up to 800 litres, it must not exceed 15 seconds - for larger chambers, it must not exceed 30 seconds

 

Effective sterilisation must be verified using chemical (ISO 11140-1:2014) 4 and biological indicators (general requirements: ISO 1138-1: 2017 5, most heat: ISO 11138-3:2017) 6. Of these different indicators, chemical indicators verify exposure to sterilisation conditions and biological indicators (e.g., Geobacillus stearothermophilus) confirm lethality.

Summary

 

This article has looked at the operational features of autoclaves and the physical control requirements for achieving sterilisation. In summary, autoclaves work by:

 

1.     Removing air

2.     Introducing saturated steam under pressure

3.     Achieving high temperatures (121–134 °C)

4.     Holding for a validated time

5.     Cooling and drying

 

The lethal mechanism for killing microorganisms is protein denaturation caused by moist heat, delivered via condensing steam.

 

There are some limitations with autoclaves in that certain materials cannot tolerate moisture or high temperatures. Autoclaves can also go wrong, failing to effectively sterilise. This can relate to air removal failures, wet loads post-autoclave operation and inadequate load configuration or appropriate validation to prove that more complex or bulky loads can be satisfactorily sterilised (often such items require extended exposure).

 

References

 

1.    ISO 17665:2024 Sterilization of health care products — Moist heat — Requirements for the development, validation and routine control of a sterilization process for medical devices


2.    EN 13060:2025 - Sterilizers for medical purposes. Small steam sterilizers. Requirements and testing


3.    EN 285:2015+A1:2021 - Sterilization: Steam sterilizers - Large sterilizers


4.    ISO 11140-1:2014 Sterilization of health care products — Chemical indicators, Part 1: General requirements


5.    ISO 11138-1:2017: General requirements for all biological indicators


6.    ISO 11138-3:2017: BIs for moist heat (steam) sterilization

 

Need to contact us? Get help or ask us a question