BY DR TIM SANDLE | PHARMACEUTICAL MICROBIOLOGY AND CONTAMINATION CONTROL EXPERT
31st March
Sterilisation by moist heat has a wide application in pharmaceuticals and healthcare. Examples of application include sterilising glassware, tools, stoppers, filters, gloves, vials, medical, laboratory and pharmaceutical items.
Autoclaves are steam based sterilisation devices that use moist heat under pressure to kill microorganisms, including highly resistant bacterial spores. Their operation is rooted in physics - raising pressure increases the boiling point of water, allowing steam to reach temperatures far above 100 °C (typically 121–134 °C), which is required for enhanced microbial kill and therefore for sterilisation.
This article examines the basic operational principles and practicalities, as required for a pharmaceutical, healthcare, medical device or food setting.
Water normally boils at 100°C at normal atmospheric pressure. However, if pressure is increased sufficiently (e.g., +1 bar / 15 psi /100 kPa), water boils at ~12 °C. This elevated temperature creates pressurised saturated steam, which is lethal to most microorganisms, including bacterial spores likely to be found in pharmaceutical facilities.
In terms of the application, steam is far more efficient at transferring heat than dry air because condensation releases large amounts of latent heat directly onto surfaces, as set out in ISO 17665:2024 1

There are several factors that make an autoclave effective in terms of achieving microbial kill robustly and within a sufficient period of time, whilst avoiding damaging the items to be sterilised. These factors are:
There are different types of autoclaves, based on their operating principles. The most common to pharmaceuticals and healthcare facilities are:
These autoclaves differ primarily in their sterilisation capabilities and air removal methods. The classes are defined by the European Standard EN 13060: 2025 2 (‘Small steam sterilizers’). In summary:
| Class | Method | Load Type | Best Use Case |
| B | Fractionated vacuum | All (wrapped, porous, hollow) | Hospitals, complex dental tools |
| S | Partial/no vacuum | Specific (manufacturer defined) | Basic wrapped, solid tools |
| N | Gravity displacement | Simple (unwrapped, solid) | Basic, non-hollow tools |
The standard EN 285: 2014 3 applies to large steam sterilisers (≥ 54 litres or a full-sized sterilisation unit). The standard itself does not classify machines into types (like ‘Class B’), whereas it defines the performance, design and testing requirements for these large, usually pre-vacuum, autoclaves (in essence Class B).
While there are differences with autoclave design, there are some common factors which apply. These are set out below:

Figure 1: A typical autoclave chart showing temperature and pressure rising through conditioning, plateauing during sterilisation and dropping during exhaust (image designed by Tim Sandle)
(1) Loading
(2) Chamber sealing
(3) Air removal (critical step)
(4) Steam admission and heating
(5) Sterilisation hold (exposure phase)
Depending on load type and mass, common validated conditions for the hold stage are:
(6) Exhaust / depressurisation
(7) Drying
Under EN 285, the important performance and technical requirements are:

Figure 2: Steam quality (designed by Tim Sandle)
The validation and testing of autoclaves require specific, rigorous tests to be conducted to ensure sterilisation efficacy, including:
Effective sterilisation must be verified using chemical (ISO 11140-1:2014) 4 and biological indicators (general requirements: ISO 1138-1: 2017 5, most heat: ISO 11138-3:2017) 6. Of these different indicators, chemical indicators verify exposure to sterilisation conditions and biological indicators (e.g., Geobacillus stearothermophilus) confirm lethality.
This article has looked at the operational features of autoclaves and the physical control requirements for achieving sterilisation. In summary, autoclaves work by:
1. Removing air
2. Introducing saturated steam under pressure
3. Achieving high temperatures (121–134 °C)
4. Holding for a validated time
5. Cooling and drying
The lethal mechanism for killing microorganisms is protein denaturation caused by moist heat, delivered via condensing steam.
There are some limitations with autoclaves in that certain materials cannot tolerate moisture or high temperatures. Autoclaves can also go wrong, failing to effectively sterilise. This can relate to air removal failures, wet loads post-autoclave operation and inadequate load configuration or appropriate validation to prove that more complex or bulky loads can be satisfactorily sterilised (often such items require extended exposure).
1. ISO 17665:2024 Sterilization of health care products — Moist heat — Requirements for the development, validation and routine control of a sterilization process for medical devices
2. EN 13060:2025 - Sterilizers for medical purposes. Small steam sterilizers. Requirements and testing
3. EN 285:2015+A1:2021 - Sterilization: Steam sterilizers - Large sterilizers
4. ISO 11140-1:2014 Sterilization of health care products — Chemical indicators, Part 1: General requirements
5. ISO 11138-1:2017: General requirements for all biological indicators
6. ISO 11138-3:2017: BIs for moist heat (steam) sterilization